Â
The YSAV50A ventilator is an electrically controlled pneumatic ventilator integrating such functions as time, volume cycling, pressure limit, etc. It is mainly intended for providing ventilation support to a critically ill patient during the life threatening phase and ensuring the going-through of the dangerous period by the patient and smooth treatment of primary diseases for recovery. Also it provides an alternation in case of irreversible lesions in respiratory muscles or irreversible damage to upper airway to maintain the respiratory function of the patient, and also provides ventilation assistance for the patient during the recovery from a disease or operation. ALL MANIN UNIT HAS PASS THE CE The automatic conversion for standby power while AC power outage, keep working not less than 1 hour TECHNIQUES SPECIFICATIONS Physical specifications
Display
LED Display
Mode:
pneumatically powered and electrically controlled system
Trolley:
Fitted with 4Â nos anti-static rubber castors; two of which are lockable for braking and easy maneuverability with foot operated brake provisions
Technical Specifications:
Gas requirement:
Medical oxygen with a pressure ranging from 280~600kpa
safety valve
<12.5 kPa
Respiratory Rate
6~60 times/ min
Oxygen concentration output
<50%
minute volume
≥18L
modes of ventilation
A/C, IPPV, SIPPV, IMV, SIMV, PEET. MANUAL, SIGH
PEEP:
0.1 ~Â 1.0 kPa
Ptr:
-4 ~Â 10 hPa
IMV frequency:
1~12 bpm
I/E Ratio:
1:1.5; 1:2.0; 1:2.5; 1:3.0
Tidal Volume
Not less than 50~ 1200ml
Inspiratory Plateau:
0~1s
Maximum Safety Pressure
≤ 6.0kPa
Sigh:
one deep breath per 80 controlled respirations, the inspiration time is 1.5 times of the setting point
Pressure limit range:
0 ~Â 6.0 kPa
Airway pressure alarm: Audible and visual and with yellow and red color indicating
Lower: 0.9 kPa ~Â 5.4 kPa;Â Â Â Upper: 0.5Â Â Â kPa
±0.2 kPa
Ventilation Volume alarm:
≥ 18L/ MIN
Power Supply Alarm
Ac/dc power supply are after failing to send out alarm immediately
Alarm Time: keep >120sThe airway pressure continues to be higher than 15 hPa ±1 hPa for 15s±1s, then the machine will raise an audible alarm, the pressure will be displayed in red and the continuous high pressure red alarm wording is displayed on the screen of the anesthetic respirator. Operating conditions
Ambient temperature:
10 ~Â 40oC
Relative humidity:
no higher than 80 %
Atmospheric pressure:
860 hPa ~Â 1060 hPa
Power requirement:
100-120 Vac, 50/60 Hz;Â
Attention: the AC power supply used for the ICU ventilator must be well   grounded. Storage
Ambient temperature:
-15oC ~Â +50oC
Relative humidity:
no higher than 95 %
Atmospheric pressure:
86 kPa ~Â 106 kPa.
Â
It should be stored in a room without corrosive gas and well ventilated
Package
packaging box
comply with the requirement of GB/T 15464
Â
Between the packaging box and product, soft material with appropriate thickness provided to prevent loosening and mutual friction during transportation
Â
Moisture protection and rain protection to ensure that product is protected from natural damage.
Safty &Â Alarm
Gas source pressure
Immediate alarm; "high level alarm" indicator flickers raising an audible alarm.
Airway Pressure Alarm
The airway pressure exceeds 15hpa. The maximum delay is no longer than 17s. The "high level alarm" indicator flickers raising an audible alarm.
Tidal Volume Alarm
Lower than 50ml, The alarm will be raised after one breathing cycle.
Ventilation Volume Alarm
Lower: 0~12L/Min;Â Â Â Up: 18L/ Min
Battery Voltage Alarm
It alrms when the battery voltage is lower than 10.5V.
Power On Self Test
Self test before working, assure it's working normally.
STANDARD CONFIGURATIONS
QTY
NAME
1 set
Main unit
1 set
Oxygen tube
2 set
Silicone Mask
1 pc
Power line
2 pcs
Threaded pipe
2 pcs
Standard Pipe
1 set
Oxygen Reducer
1 set
Humidifier
1 pc
Battery
1 set
flow sensor
2 pcs
A leather bag (Â Blue)
1 set
Tools with the machine
1 set
User Manual (English Version)
We are a company aiming to make one stop shopping for our customers. We equiped hospital
operation room with many kind of items like operation table, orthopedic drill and saw, ECG,
defibrillator etc.
We listed some hospital operation room items basic items for your reference as below:
operation room
anesthesia machine
ventilator operating lamp operation table Pendant medical drill and saw ECG patient monitor Defibrillator infusion pump Electrocoagulation oxygen generator electric suction device autoclave sterilizer
Â
Medical Surgical Electrically Respirator Ventilator
Model NO.: YSAV50A
Respiratory Rate: 6~60 Times/ Min
Oxygen Concentration Output: <50%
Peep: 0.1 ~ 1.0 Kpa
Imv Frequency: 1~12 Bpm
Ptr: -4 ~ 10 Hpa
I/E Ratio: 1:1.5; 1:2.0; 1:2.5; 1:3.0
Trademark: YSENMED
Transport Package: Wooden Case
Specification: CE certificate
Origin: China
HS Code: 9018909000
YSAV50A Medical Surgical Electrically Respirator VentilatorÂ