Does the Shanghai incubator drying box need to be metered?

Is the incubator of the incubator required to be metered? Small incubators generally do not require metering, and only thermometers are required if necessary. The incubator and drying oven should be calibrated according to JJF1101-2003 for temperature deviation, temperature uniformity and temperature fluctuation. You only have one calibrated thermometer that cannot calibrate the incubator and the drying oven.

When the volume of the incubator is more than 2 cubic meters, 15 temperature probes should be used at the same time. When the volume of the incubator is less than or equal to 2 cubic meters, the temperature of the specified position in the tank should be measured simultaneously with 9 temperature probes at the same time for 15 times. Deviation, temperature uniformity and temperature fluctuations.