1. The fuse is blown
This kind of fault occurs more and there must be a short circuit. After the power is cut off, the electric heating wire of each electric heating pipe can be used to measure whether the electric heating wire is grounded. If it is grounded, the two ends of the electric heating wire can be removed. This can be used as an emergency measure. The distiller is still available, but the power is reduced, and the water output is corresponding. Reduce or replace it. For example, the 20 liter/hour electric distilled water is 6 sets of 3 sets, which are used in parallel. Sometimes the electric heating wire is not grounded when the power is off, and the grounding is burned when the power is turned on for a while. At this time, you can check whether the scale is thick, if it is thick, it can be removed; if it is not thick, it may be caused by the thermal expansion of the electric heating wire of a certain electric heating tube.
2. Close the power switch and make a sound and spark
Closing the power switch has a sound and spark, indicating that the electrical circuit has a short circuit or contact, causing the current to be too large. The cause of excessive current is not only the grounding of the heating wire in the electric heating tube, but also the leakage between the electric heating tube and the evaporation pot, causing a short circuit at the power connector, or the ignition of the power connector due to moisture, resulting in the resistance of the wiring insulation board. Become smaller.
3. Reduced amount of distilled water
As the use time increases, the scale of the electric heating pipe also gradually increases, the heat of the electric heating pipe is not easily transmitted, and the amount of distilled water is gradually reduced, and the scale can be removed by physical or chemical methods. If the amount of water is significantly reduced, there must be a heating wire that is blown, causing a decrease in power and a decrease in the amount of water. A multimeter can be used to detect the burnt electric heating tube and replace it. After replacement, the water should be drained to observe whether it is leaking or seeping water. It can be energized only after there is no water seepage.